Anasazi Instruments

FT-NMR Instruments for Education and Industry

Optimizing Lineshape 60MHz

Instructions for optimizing the lineshape for the Eft NMR.

(Note: The PNMR instructions are printed in blue.  NUTS commands are printed in red.)

1) For spinning rate and referencingwater sample - spinning

  • run the prep routine (type: prep <Enter> and follow onscreen prompts)
  • you’ll need to slow the spinning at one point and then speed it up later.
  • Make sure the sample is spinning at a constant speed (4-6 squares apart)
  • shims X, Z, C and Y to minimize spinning sidebands and improve spinning lineshape
  • Also sets the correct spin speed for later data acquisition.
  • takes about 5 minutes

2) For lineshape optimization

  • water sample - spinning
  • run the shim2 routine (type: shim2 <Enter> and follow onscreen prompts)
  • shims 4th, 3rd, C and Y to improve spinning lineshape
  • takes about 15-20 minutes

3) Reduce spinning sidebands further

  • water sample - spinning
  • run the suprep routine (type: suprep <Enter> and follow onscreen prompts)
  • you’ll need to slow the spinning at one point and then speed it up later.
  • Make sure the sample is spinning at a constant speed (4-6 squares)
  • shims XY, X, ZY, Z, C and Y to eliminate spinning sidebands and improve spinning lineshape
  • Also sets the correct spin speed for later data acquisition.
  • takes about 20 minutes

NOTE: Sometimes, you’ll need to repeat steps 2 and 3

4) OPTIONAL: For final lineshape optimization

  • 5% ethyl methacrylate or 5% ethyl benzene or 5% ethyl trans-crotonate – spinning
  • run the shim3 routine (type: shim3 <Enter> and use a 5 second relaxation delay)
  • shims 3rd, C and Y to improve spinning lineshape for organic samples
  • takes about 15 minutes

5) Calibrate the pulse widths and decoupler

  • 98% n-propylbenzoate sample – spinning
  • Shim the sample (type: shim <Enter> and use RD=2)
  • Set the reference(type: zgha2fozgha2, etc…)
  • Go to the TST> prompt (type: nu tst <Enter> )
  • Run the full Calibration (type: chcal <Enter> <Enter> )
  • Call to make sure the numbers make sense (call us if they don’t at 866-494-9369)
  • Run Calibration again (type: chcal <Enter> <Enter> )

6) Check the calibration

  • 98% n-propylbenzoate sample – spinning
  • Go to the C13> prompt (type: nu C13 <Enter> )
  • Run a DEPT experiment (type: dept <Enter> )
  • Process the DEPT. (in NUTS type: <CTRL>+F11 )